Multiple-sprayer tandem mass spectrometry with parallel high flow extraction and parallel separation for high-throughput quantitation in biological fluids.
Authors:Y Deng  H Zeng  S E Unger  J T Wu
Institution:Stine-Haskell Research Center, DuPont Pharmaceuticals Company, PO Box 30, Newark, DE 19714, USA.
Abstract:In this work, an on-line four-column high-flow parallel extraction device was set up and interfaced to a four-column parallel separation system with a four-sprayer tandem mass spectrometric detector. This system allowed for the direct and simultaneous sampling, extraction, separation, and detection of four samples in biological fluids, resulting in significantly increased throughput. The performance of this system was evaluated using rat plasma samples spiked with methotrexate as a test compound. The results showed that the relative standard deviation for responses across the four parallel channels was less than 10% for a concentration range from 10 to 2500 nM, resulting in comparable sensitivity (slope of curve) for the analyte on all channels. A ruggedness test was performed using a mixed set of new and used extraction columns and the relative standard deviation in analyte response was found to be below 15%. The inter-channel crosstalk was measured to be about 0.1%. A comparison between the data generated on this system and on a conventional system for a rat pharmacokinetic study showed no more than 20% difference in measured concentrations. This parallel system is a viable tool for high-throughput bioanalysis in a drug discovery environment.