首页 | 本学科首页   官方微博 | 高级检索  
     


Frequency response of the glucose sensor based on immobilized glucose oxidase membrane
Authors:Gondo  Shinichiro  Kim  Chulsoon  Morishita  Michio  Kawakami  Mitsuyasu
Affiliation:1.Department of Electronic Materials Engineering, Fukuoka Institute of Technology, Wajirohigashi, 811-02, Higashiku, Fukuoka, Japan
;
Abstract:

Frequency response of the glucose sensor based on the immobilized glucose oxidase membrane was investigated experimentally by giving the sinusoidal change of glucose concentration to the glucose sensor and observing its output signal. Observed values of gains and phase lags of the frequency response of the glucose sensor followed the frequency response model of the first-order with dead time; The time constant and also the dead time were estimated and found to decrease as the amount of enzyme immobilized in the membrane increased and the thickness of the membrane decreased.

Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号