首页 | 本学科首页   官方微博 | 高级检索  
     


A Comparison of Four Different Electrode Matrices on the Performance of Amperometric Hydrogen Peroxide (Bio)Sensors
Authors:Kubilay Tunca  Funda Öztürk  Pınar Esra Erden
Affiliation:1. Department of Chemistry, Faculty of Science and Arts, Tekirdağ Namık Kemal University, 59030 Tekirdağ, Turkey;2. Department of Chemistry, Polatlı Faculty of Science and Letters, Ankara Hacı Bayram Veli University, 06900 Ankara, Turkey
Abstract:A comparison of the analytical performances of four different (bio)sensor designs in H2O2 determination is discussed. The (bio)sensor designs developed were based on the use of (i) multiwalled carbon nanotubes (MWCNT), zinc oxide nanoparticles (ZnONP), prussian blue (PB); (ii) MWCNT, ZnONP, PB and ionic liquid (IL); (iii) MWCNT, ZnONP and horseradish peroxidase (HRP) and (iv) MWCNT, ZnONP, HRP and IL modified glassy carbon electrode (GCE). A performance comparison of (bio)sensors showed that the one based on HRP/IL-MWCNT-ZnONP/GCE showed the best analytical characteristics with a linear dynamic range of 9.99×10−8–7.55×10−4 M, detection limit of 1.37×10−8 M and sensitivity of 17.00 μA mM−1.
Keywords:Amperometric (bio)sensor  hydrogen peroxide  carbon nanotubes  zinc oxide nanoparticles  ionic liquid  prussian blue