Short communication: A simple and accurate method of measuring the zeta-potential of microfluidic channels
Authors:Raúl Fernández-Mateo  Pablo García-Sánchez  Víctor Calero  Antonio Ramos  Hywel Morgan
Institution:1. School of Electronics and Computer Science, University of Southampton, Southampton, United Kingdom;2. Depto. Electrónica y Electromagnetismo, Facultad de Física, Universidad de Sevilla, Sevilla, Spain
Abstract:We describe an improved method for determining the electroosmotic mobility and zeta potential of surfaces based on a current-monitoring method. This technique eliminates the requirement for measurements of channel dimensions and sample conductivities, leading to a simple high precision measurement. The zeta potential of PDMS is measured for native surfaces and surfaces treated with a nonionic surfactant in low-conductivity electrolytes.
Keywords:current-monitoring method  electroosmosis  PDMS  zeta-potential