首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Semi-automatic catalytic titration of some aminopoly-carboxylic acids,copper(II), and cobalt(II)
Authors:M M Timotheou-Potamia  M A Koupparis  T P Hadjiioannou
Institution:(1) Laboratory of Analytical Chemistry, University of Athens, Athens, Greece
Abstract:Summary A semi-automatic potentiometric method is described for the catalytic titrimetric determination ofmgrg amounts of some aminopolycarboxylic acids. The method is based on their inhibitory effect on the copper(II)-catalyzed periodate-thiosulfate reaction. Amounts of EDTA in the 0.7–600mgrg range (10–6–8×10–5 M), of DCTA in the 7–7000mgrg range (10–6–10–3 M), of EGTA in the 0.8–800mgrg range (10–7–10–4 M), and of DTPA in the 4–800mgrg range (5×10–7–10–4 M) were determined with average relative errors and coefficient of variation of about 0.4–1%. The method has also been used for the indirect catalytic titrimetric determination ofmgrg amounts of Cu2+ and Co2+ ions with about the before mentioned accuracy and precision.
Halbautomatische katalytische Titration einiger Aminopolycarbonsäuren sowie von Cu(II) und Co(II)
Zusammenfassung Eine halbautomatische potentiometrische Methode zur katalytischen Maßanalyse von Mikrogrammengen einiger Aminopolycarbonsäuren wurde angegeben. Sie beruht auf dem Hinderungseffekt gegenüber der Cu(II)katalysierten Perjodat-Thiosulfatreaktion. EDTA in Mengen von 0,7–600mgrg, DCTA in Mengen von 7–7000mgrg, EGTA in Mengen von 0,8–800mgrg und DTPA in Mengen von 4–800mgrg wurden mit einem mittleren relativen Fehler von etwa 0,4–1% bestimmt. Das Verfahren wurde auch zur Bestimmung von Cu(II) und Co(II) mit der angeführten Genauigkeit verwendet.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号