首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Correction of accurate mass measurement for target compound verification by quadrupole time-of-flight mass spectrometry
Authors:Harald?C?K?feler  Email author" target="_blank">Michael?L?GrossEmail author
Institution:Department of Chemistry, Washington University, St. Louis, Missouri 63130, USA.
Abstract:The aim of this work is to evaluate quadrupole/time-of-flight (QTOF) mass spectrometry for simultaneous measurement of accurate mass and quantification of a target by using a stable isotopically labeled internal standard. Mixtures of caffeine and (13)C(3)-caffeine (internal standard) at different concentration ratios were analyzed by capillary HPLC/QTOF. A calibration plot for quantification is linear over a factor of 20. Without invoking any correction scheme, the mass accuracy seriously degraded when the ratio of the mass standard to the test compound was not unity. The accuracy could be restored to approximately 2 ppm by using a quadratic function to correct the measured mass as a function of the measured signal ratio of target and internal calibrant.
Keywords:
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏