Method of Separating the Sensitive Volume of Calorimetric Cells in a Differential Titration Calorimeter |
| |
Authors: | Kotelnikov G V Moiseyeva S P Mezhburd E V Krayev V P |
| |
Institution: | (1) Institute for Biological Instrumentation, Russian Academy of Sciences, 142292 Pushchino Moscow Region, Russia |
| |
Abstract: | A scientific and technical solution of capillary calorimetric unit design is described based on a method of separating the
sensitive volume of calorimetric cells with a heat-conducting bridge in a differential titration calorimeter.
The design of the calorimetric cells as thin capillary tubes allowed to minimize the dimension of the heat-conducting bridges
and to separate with high accuracy the sensitive volume of the calorimetric cells which is of 78.5 μl. Due to high calorimeter
power sensitivity (10 nW) a modern level of requirements for measuring minimum specific heat of processes is provided.
The adopted design of the calorimetric cells permits to possess the cells located horizontally and to simplify the solution
of the problem of equalizing components concentration along the volume of the calorimetric cells.
This revised version was published online in July 2006 with corrections to the Cover Date. |
| |
Keywords: | differential titration calorimeter heat-conducting bridge sensitive volumes of calorimetric cells |
本文献已被 SpringerLink 等数据库收录! |