首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Stripping Voltammetric Determination of Methyl Parathion at Activated Carbon Nanopowder Modified Electrode
Authors:Margarita Stoytcheva  Roumen Zlatev  Zdravka Velkova  Velizar Gochev  Gisela Montero  Benjamín Valdez  Mario Curiel
Institution:1. Universidad Autónoma de Baja California, Instituto de Ingeniería, Mexicali, México;2. Medical University of Plovdiv, Faculty of Pharmacy, Dep. Chemical Sciences, Plovdiv, Bulgaria;3. Plovdiv University “P. Hilendarski”, Faculty of Biology, Dep. Biochemistry and Microbiology, Plovdiv, Bulgaria
Abstract:An activated carbon nanopowder modified glassy carbon electrode (AC-GCE) was constructed for the sensitive determination of methyl parathion by adsorptive differential pulse anodic stripping voltammetry. The simple and rapid modification procedure included only drop-coating the electrode surface with a laponite stabilized activated carbon nanopowder suspension and drying. The modifier high adsorption ability, combined with its large electroactive surface area allowed a 30-fold signal increase to be achieved, compared to bare GCE. Under optimized experimental conditions (activated carbon to laponite ratio, pH and accumulation time), the AC-GCE exhibited a linear response to methyl parathion in two concentration ranges: from 0.01 μmol L?1 to 1 μmol L?1 and from 1 μmol L?1 to 6 μmol L?1. The LOD of 2.5 nmol L?1 (S/N=3) achieved fitted with regulatory norms. It was demonstrated that the as-prepared AC-GCE is suitable for routine real samples analysis.
Keywords:Methyl parathion  Solid-phase extraction  Activated carbon nanopowder  Stripping voltammetry
设为首页 | 免责声明 | 关于勤云 | 加入收藏