首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Measurement of the radon exhalation rate from the medium surface by tracing the radon concentration
Authors:Yanliang Tan  Detao Xiao
Institution:1. Department of Physics and Electronics, Hengyang Normal University, Huangbai Road 165#, Hengyang, 421008, Hunan, China
2. School of Nuclear Science & Technology, University of South China, Hengyang, 421001, Hunan, China
Abstract:The paper will present a method based on the accumulation chamber technique for measuring of radon exhalation from the medium surface. A radon monitor traces the change of radon concentration in the accumulation chamber, and then the radon exhalation can be obtained accurately through linear fit. Based on our recent experiments, the radon exhalation rate from the medium surface obtained from this method is in good agreement with the actual exhalation rate of our simulation facility. This method is superior to the competition method which obtains the radon exhalation through the exponential fit by an external PC-system. The calculation for the exponential fit is very easy by computer and related software. However, for portable instruments, the single chip microcomputer can’t calculate the exponential fit rapidly. Thus, this method is usable for developing the new portable instrument to classify building materials, etc.
Keywords:
本文献已被 SpringerLink 等数据库收录!