首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Multiplexed proteomic sample preconcentration device using surface-patterned ion-selective membrane
Authors:Lee Jeong Hoon  Song Yong-Ak  Han Jongyoon
Institution:Department of Electrical Engineering and Computer Science, Department of Biological Engineering, Massachusetts Institute of Technology, 77 Massachusetts Avenue, Cambridge, Massachusetts 02139, USA.
Abstract:In this paper, we report a new method of fabricating a high-throughput protein preconcentrator in poly(dimethylsiloxane) (PDMS) microfluidic chip format. We print a submicron thick ion-selective membrane on the glass substrate by using standard patterning techniques. By simply plasma-bonding a PDMS microfluidic device on top of the printed glass substrate, we can integrate the ion-selective membrane into the device and rapidly prototype a PDMS preconcentrator without complicated microfabrication and cumbersome integration processes. The PDMS preconcentrator shows a concentration factor as high as approximately 10(4) in 5 min. This printing method even allows fabricating a parallel array of preconcentrators to increase the concentrated sample volume, which can facilitate an integration of our microfluidic preconcentrator chip as a signal enhancing tool to various detectors such as a mass spectrometer.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏