Validation of a confirmatory method for the determination of residues of four nitrofurans in egg by liquid chromatography-tandem mass spectrometry with the software InterVal |
| |
Authors: | Bock C Stachel C Gowik P |
| |
Affiliation: | Federal Office of Consumer Protection and Food Safety (BVL), EU- and National Reference Laboratory (CRL/NRL), Diedersdorfer Weg 1, 12277 Berlin, Germany. |
| |
Abstract: | A method for the detection and determination of nitrofuran derivatives in egg by liquid chromatography-tandem mass spectrometry (LC-MS/MS) was validated with the software InterVal and can be applied for the confirmation of nitrofuran metabolites in fresh or lyophilised eggs. The validation study comprises variations in operator, storage condition, breeding, equipment and duration of sample preparation. A comprehensive overview of the robustness of the method is obtained by analysing eight samples at six concentration levels. First results of short- and medium-term investigations for stability of analytes in solution show that standard solutions of nitrofuran metabolites are stable for at least 1 year when stored at +4 degrees C in the dark. The decision limit CCalpha expressed for the underivatised metabolite is 0.05 microg kg(-1) for 3-amino-5-methyl-morpholino-2-oxazolidinone, 0.03 microg kg(-1) for 3-amino-2-oxazolidinone, 0.20 microg kg(-1) for semicarbazide and 0.22 microg kg(-1) for 1-amino-hydantoin. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|