首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Automated measurement and classification of pulmonary blood-flow velocity patterns using phase-contrast MRI and correlation analysis
Authors:van Amerom Joshua F P  Kellenberger Christian J  Yoo Shi-Joon  Macgowan Christopher K
Institution:Department of Medical Imaging, University of Toronto, and Department of Diagnostic Imaging, Hospital for Sick Children, Toronto, Ontario, Canada.
Abstract:An automated method was evaluated to detect blood flow in small pulmonary arteries and classify each as artery or vein, based on a temporal correlation analysis of their blood-flow velocity patterns. The method was evaluated using velocity-sensitive phase-contrast magnetic resonance data collected in vitro with a pulsatile flow phantom and in vivo in 11 human volunteers. The accuracy of the method was validated in vitro, which showed relative velocity errors of 12% at low spatial resolution (four voxels per diameter), but was reduced to 5% at increased spatial resolution (16 voxels per diameter). The performance of the method was evaluated in vivo according to its reproducibility and agreement with manual velocity measurements by an experienced radiologist. In all volunteers, the correlation analysis was able to detect and segment peripheral pulmonary vessels and distinguish arterial from venous velocity patterns. The intrasubject variability of repeated measurements was approximately 10% of peak velocity, or 2.8 cm/s root-mean-variance, demonstrating the high reproducibility of the method. Excellent agreement was obtained between the correlation analysis and radiologist measurements of pulmonary velocities, with a correlation of R2=0.98 (P<.001) and a slope of 0.99+/-0.01.
Keywords:Correlation analysis  Pulmonary blood flow  Phase contrast  Pulmonary hypertension  Congenital heart disease
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号