New possibilities of recently constructed high-temperature solution calorimeter |
| |
Authors: | Tomasz Maciąg Krzysztof Rzyman |
| |
Affiliation: | 1. Department of Metallurgy, Silesian University of Technology, Krasińskiego 8, 40-019, Katowice, Poland
|
| |
Abstract: | The calorimeter presented in this paper is constructed under supervision of authors. The most substantial innovation of construction of the device was introduction of intermediate container with ten sockets. As a result the calorimeter is used for determination of enthalpy of formation for alloys in a wide range of temperature (up to 1,400 K). As an example results from alloys from the Ni–Al–Cr system are presented, they are aimed to determine directly γ′ / γ′ + γ / γ phase boundaries with use of this “in situ” calorimetric method. Precise indication of phase boundaries is crucial for determination of phase diagrams. |
| |