The efficiency of immobilised glutamate oxidase decreases with surface enzyme loading: an electrostatic effect, and reversal by a polycation significantly enhances biosensor sensitivity |
| |
Authors: | McMahon Colm P Rocchitta Gaia Serra Pier A Kirwan Sarah M Lowry John P O'Neill Robert D |
| |
Affiliation: | UCD School of Chemistry and Chemical Biology, University College Dublin, Belfield, Dublin 4, Ireland. |
| |
Abstract: | The apparent Michaelis constant, K(M), for glutamate oxidase (GluOx) immobilised on Pt electrodes increased systematically with enzyme loading. The effect was due, at least in part, to electrostatic repulsion between neighbouring oxidase molecules and the anionic substrate, glutamate (Glu). This understanding has allowed us to increase the Glu sensitivity of GluOx-based amperometric biosensors in the linear response region (100+/-11 nA cm(-2)microM(-1) at pH 7.4; SD, n=23) by incorporating a polycation (polyethyleneimine, PEI) to counterbalance the polyanionic protein. Differences in the behaviour of glucose biosensors of a similar configuration highlight a limitation of using glucose oxidase as a model enzyme in biosensor design. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|