首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Liquid chromatographic tandem mass spectrometric determination of trandolapril in human plasma
Authors:Constantinos Pistos  Maria Koutsopoulou  Irene Panderi
Institution:

aIndependent Research and Laboratory Solutions (ILS), 240 Klisthenous Street, 153 44 Gerakas, Athens, Greece

bSchool of Pharmacy, Division of Pharmaceutical Chemistry, University of Athens, Panepistimiopolis, Zografou 157 71, Athens, Greece

Abstract:A liquid chromatographic tandem mass spectrometric method for the determination of trandolapril in human plasma has been developed and fully validated. The article describes, in detail, the bioanalytical procedure and summarizes the validation results obtained. The samples were extracted using HLB Oasis solid-phase extraction cartridges. The chromatographic separation was performed on X-Terra C8 MS column (150 mm × 4.6 mm i.d., 5 μm particle size) using a mobile phase consisting of acetic acid 20 mM and triethylamine 4.3 mM/acetonitrile (40:60 (v/v)), pumped isocratically at 0.35 ml/min.

The analytes were detected using a micromass quattro micro triple quadrupole mass spectrometer with positive electrospray ionization in multiple reaction-monitoring (MRM) mode. Tandem mass spectrometric detection enabled the quantitation of trandolapril down to 2.0 ng/ml. Calibration graphs were linear (r better than 0.996, n = 9) in the concentration ranges 2.0–750 ng/ml and the intra- and inter-day R.S.D. values were less than 3.83 and 3.86% for trandolapril.

Keywords:Trandolapril  Human plasma  Tandem mass spectrometry
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号