首页 | 本学科首页   官方微博 | 高级检索  
     

电荷注入检测器电感耦合等离子体光谱仪测定非金属元素的分析性能
引用本文:辛仁轩,王建晨. 电荷注入检测器电感耦合等离子体光谱仪测定非金属元素的分析性能[J]. 分析化学, 2002, 30(11): 1375-1378
作者姓名:辛仁轩  王建晨
作者单位:清华大学核能技术设计研究院,北京,100084;清华大学核能技术设计研究院,北京,100084
摘    要:研究了电荷注入检测器ICP光谱仪测定非金属元素的分析性能。以S、P、As、Se为代表的非金属元素在170-800nm波段内最灵敏的谱线均处于175-200nm远紫外区内。在此区内等离子体有很低的光谱背景发射和良好的谱线测量和背景测量的光度精度。标准曲线线性动态范围在4个数量级。给出了As,Se,S,P主要分析线的灵敏度、线背比、背景等效浓度及检出限。在纯水溶液中的检出限分别为:As 189.142nm 0.003mg/L,P213.618nm 0.005mg/L,S 180.731nm 0.01mg/L,Se 196.090nm 0.009mg/L.

关 键 词:等离子体发射光谱  非金属元素          电荷注入检测器

Analytical Performance of Inductively Coupled Plasma Atomic Emission Spectrometer with Charge Injection Device for the Determination of Non-metal Elements
Xin Renxuan ,Wang Jianchen. Analytical Performance of Inductively Coupled Plasma Atomic Emission Spectrometer with Charge Injection Device for the Determination of Non-metal Elements[J]. Chinese Journal of Analytical Chemistry, 2002, 30(11): 1375-1378
Authors:Xin Renxuan   Wang Jianchen
Affiliation:Xin Renxuan *,Wang Jianchen
Abstract:Analytical performance for the detection of non-metal elements by inductively coupled plasma atomic emission spectrometer (ICP-AES) with charge injection device(CID) was investigated. The experimental results showed that the photometric precision of determination of spectral lines and background were excellent for S, P, As and Se. The linear range of calibration curves was 0.03 mg/L to 1000 mg/L for As, and 0.1 mg/L to 1000 mg/L for P, 0.3 mg/L to 1000 mg/L for S, and 1.0 mg/L to 1000 mg/L for Se. The detection limits were 0.003 mg/L for As, 0.005 mg/L for P, 0.01 mg/L for S, 0.009 mg/L for Se. The information about the determination of non-metals such as sensitivity, line-to-background ratio, equivalent concentration of background was examined.
Keywords:Inductively coupled plasma-atomic emission spectrometry   non-metal elements   sulfur   phosphorus   arsenic   selenium   charge injection device
本文献已被 CNKI 维普 万方数据 等数据库收录!