首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Spectrofluorimetric flow injection determination of trace amounts of periodate
Authors:Ensafi A A  Dehahgi G B
Institution:College of Chemistry, Isfahan University of Technology, Iran. ensafi@cc.iut.ac.ir
Abstract:A sensitive, rapid and selective procedure is proposed for the flow injection determinations of periodate by spectrofluorometric detection. The method is based on the reaction of periodate with Alizarin Navy Blue in basic solution. The reagents and manifold variables influence on the sensitivity have been investigated and the optimum conditions are established. Periodate can be determined for the range of 0.250-5.00 microg ml(-1) with a limit of detection of 0.08 microg ml(-1), and with a sample rate of 15 +/- 2 samples h(-1). The relative standard deviations for eight replicate determination of 0.500 and 5.00 microg ml(-1) was 1.3 and 1.1%, respectively. Periodate can be determined in the presence of iodate and bromate. The proposed method was used to determination of periodate in water samples.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!