Abstract: | Analytical procedures for the measurement of testosterone by mass fragmentography (MF) using trideuterated testosterone (testosterone-19,19,19-d3) are described. For the calculation of plasma testosterone, peak height ratios were measured by MF performed on the molecular ions of the TFA derivative of testosterone (m/e 480) and testosterone-19,19,19-d3 (m/e 483). The sensitivity of the method was judged from the lower limit of detection of the mass spectrometer which was at 10 pg. For the measurement of the precision, the inter- and intra-assay coefficients of variation (C.V.) were calculated by using a pooled plasma sample; they were 3.15% and 1.79%, respectively. The specificity was investigated by the use of 5 alpha-dihydrotestosterone and the MF method was found to afford a highly selective technique. These results obtained by MF have been compared with the results obtained by a radioimmunoassay method. |