首页 | 本学科首页   官方微博 | 高级检索  
     


Colorimetric measurement of serum and urine magnesium with the Cobas Mira Analyzer
Authors:Dieter Ratge   Klaus Peter Kohse   Ernst Knoll  Hermann Wisser
Affiliation:(1) Department of Clinical Chemistry, Robert-Bosch-Krankenhaus, Auerbachstr. 110, D-7000 Stuttgart 50, Federal Republic of Germany
Abstract:Conclusion The major advantage of the xylidyl blue-1 method over kits using the calmagite reagent is the stability of the alkaline reagent mixture ready for use [5]. Furthermore, the kit benefits from fewer interferences than the calmagite procedure [3]. The kit performed well in terms of precision, sensitivity and linearity and the results correlate excellently with those by atomic absorption spectrophotometry.
Colorimetrische Bestimmung von Magnesium im Serum und Urin mit dem Cobas Mira Analysator
Keywords:
本文献已被 SpringerLink 等数据库收录!