首页 | 本学科首页   官方微博 | 高级检索  
     


Simultaneous determination of Cu, Cd and Pb in drinking-water using W-Coil AAS
Authors:Salido A  Jones B T
Affiliation:Department of Chemistry, Wake Forest University, Winston-Salem, NC 27109-7486, USA.
Abstract:An inexpensive, multi-element, W-coil atomic absorption spectrometer has been developed. Atomization occurs on W-coils extracted from commercially available slide projector bulbs. The system has minimal power requirements, 120 ACV and 15 A. A small, computer controlled CCD spectrometer is used as the detector. A multi-element Cu, Cd and Pb hollow cathode lamp is used as the source. 20 mul volumes are deposited on the coil and atomized at 6.7 A or approximately 2200 degrees C. Cu, Cd and Pb were simultaneously determined in tap water, drinking water and a quality control sample. The instrument detection limits are 0.8, 0.2 and 3.0 mug/l for Cu, Cd and Pb, respectively.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号