首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Immobilization of 8-hydroxyquinoline onto silicone tubing for the determination of trace elements in seawater using flow injection ICP-MS
Authors:Willie S N  Tekgul H  Sturgeon R E
Institution:Insititute for National Measurement Standards, National Research Council of Canada, Ottawa, Ont. K1A OR9, Canada.
Abstract:A rapid and simple on-line method is described for the preconcentration of Mn, Co, Ni, Cu, Zn, Cd and Pb from sea water using 8-hydroxyquinoline immobilized onto silicone tubing (Sil-8-HQ) via the Mannich reaction. Recoveries between 35 and 95% and limits of detection in the ppt range were obtained using a 2 m long Sil-8-HQ tube with a sample flow rate of 1.0 ml min(-1). A tube could be subjected to sample loading and elution cycles over 200 times. The capacity was 1.5 and 1.3 mug cm(-2) for Cu and Mn, respectively. Cu, Cd, Co, Pb, Mn, Zn and Ni were determined in coastal and open ocean seawater using flow injection inductively coupled plasma mass spectrometry (FI-ICP-MS). Good agreement with certified values for the certified reference materials NASS-4 and CASS-3 was demonstrated when quantitation was undertaken by the method of additions.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号