首页 | 本学科首页   官方微博 | 高级检索  
     


Microfabricated glucose biosensor with glucose oxidase entrapped in sol-gel matrix
Authors:Yang S  Lu Y  Atanossov P  Wilkins E  Long X
Affiliation:Chemical and Nuclear Engineering Department, The University of New Mexico, Albuquerque, NM 87131-1341, USA.
Abstract:A microfabricated glucose biosensor based on an amperometeric hydrogen peroxide electrode has been developed. A sol-gel layer with 5 A pore size and 2 mum thickness was used as the glucose oxidase entrapping matrix. The sol-gel matrix formed over the silicon-based sensor has good mechanical and chemical stability, and the ability to entrap a large amount of enzyme. The miniaturized electrode sensing system is composed of platinum as both working and counter electrodes and silver as a reference electrode. Nafion(R) coating was applied as the interference limiting layer. A series of technologies, such as standard photolithography, electron beam evaporation and image reverse lift-off were utilized for mass production allowing 143 electrodes to be produced at the same time. The effect of oxidable interferences was <10% of the background value of the sensor response. Calibration tests of a series of individual sensors manufactured from the same silicon wafer and dip coated in the same conditions, showed a highly reproducible response characteristics (linear range up to 500 mg dl(-1) and mean sensitivity of 0.54+/-0.14 nA mg(-1) dl(-1) (n=10)).
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号