首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of the precision and accuracy of an automated sample introduction accessory for the flame atomic absorption spectrometric measurement of calcium in serum
Authors:Epstein M S  Buehler B  Bullard M F
Affiliation:Department of Science, Mount Saint Mary's College, Emmitsburg, MD 21727-7799, USA.
Abstract:The measurement precision and accuracy that results from the use of a sample introduction peristaltic pump system for automated instrument calibration and analysis is compared to manual methods based on traditional gravimetric and volumetric dilution procedures for flame atomic absorption analyses in water and serum matrices. Whilst use of the automated system improves speed of analysis, measurement precision was found to be approximately 2-fold worse than the manual methods based on gravimetric and volumetric dilution procedures.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号