Abstract:A new electrochemical cell assembly with the combination of UV and amperometric detector (AD) based on their complementarity was described. A Nafion tubing junction was used to decouple the high voltage from the separation capillary in the rear of on-column UV detector. In this mode, the electroactive and inert compounds could be detected by UV and AD at the same time. Aromatic amines were determined with the UV and the end-column AD detection to evaluate the performances of the cell assembly. Such an improved electrochemical detector could match the capillary with different diameters. By simple adjustment of the screws, the positioning of the working electrode and the detection capillary was easily gained without microscope. It is also very easy to assemble and disassemble the working electrode when needed.