首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Calibration of the Hg chalcogenide glass membrane ion-selective electrode in seawater media
Authors:De Marco R  Shackleton J
Institution:School of Applied Chemistry, Curtin University of Technology, GPO Box U 1987, Perth, WA 6845, Australia.
Abstract:It is shown that a chalcogenide glass mercury(II) ion-selective electrode (ISE) can be calibrated in chloride-free unbuffered and saline buffered standards, displaying near-Nernstian response over 19 orders of magnitude (i.e. 10(-20) to 10(-1) M Hg(2+)). Extended ageing of the ISE in seawater induced a memory effect, causing the electrode to respond in a sub-Nernstian fashion. Electrochemical impedance spectroscopy (EIS) demonstrated that the response of the Hg(II) ISE is underpinned by a charge transfer process, and seawater matrix effects are due to electrode passivation. It is shown that standard addition ISE potentiometry may compensate for interferences caused by the seawater matrix.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏