首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of plasma heparin by micellar electrokinetic capillary chromatography
Authors:Zhou X M  Liu J W  Zhang M E  Chen S J
Affiliation:Department of Medical Laboratory, NanFang Hospital, Guangzhou 510515, People's Republic of China.
Abstract:The micellar electrokinetic capillary chromatography (MECC) method is reported for the separation of heparin, and for the possibility of direct determination of free heparin in plasma. The conditions for MECC were: pH 8.5, 25 mM sodium dodecyl sulfate (SDS), 25 mM borate buffer, with a 30 cmx50 mum ID fused-silica capillary. The sample was detected with a UV-detector at 270 nm with heparin as external standard. The recovery rate was 95.6-98.7%. This method was linear in the range 80-7000 U l(-1). The within-run and between-run relative standard deviations were lower than 3.1 and 4.5%, respectively. It is suggested that this MECC method may be used to determine blood samples containing high levels of heparin.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号