A metallic cobalt electrode for the indirect potentiometric determination of calcium and magnesium in natural waters using flow injection analysis |
| |
Authors: | Chen Z Adams M A |
| |
Affiliation: | Department of Botany, The University of Western Australia, Nedlands, WA 6907, Australia. |
| |
Abstract: | A flow injection analysis of Ca(2+) and Mg(2+) using indirect potentiometric detection in natural waters is proposed, where Ca(2+) or Mg(2+) are injected into a buffer carrier containing phosphate, resulting in the formation of Ca(3)(PO(4))(2) or Mg(3)(PO(4))(2). The consequent reduction in free phosphate in the carrier solution is detected using a metallic cobalt wire electrode. Indirect electrode response was used and the experimental conditions affecting electrode response were optimized. Responses were linear in the concentration range 5x10(-4) to 5x10(-3) M with a detection limit of 1x10(-5) M in 20 mM phosphate buffer at pH 8.0. The relative standard derivation at 1 mM of Ca(2+) and Mg(2+) were 3.9 and 3.7% (n=10), respectively. EGTA and 8-hydroxyquinoline were used as the masking agents for Ca(2+) and Mg(2+), respectively. Concentrations of Ca(2+) and Mg(2+) in natural waters were successfully determined by the proposed method. |
| |