首页 | 本学科首页   官方微博 | 高级检索  
     


Development of an amperometric biosensor based on acetylcholine esterase covalently bound to a new support material
Authors:Khayyami M  Pérez Pita M T  Peña Garcia N  Johansson G  Danielsson B  Larsson P O
Affiliation:Department of Pure and Applied Biochemistry, Centre for Chemistry and Chemical Engineering, Lund University, S-221 00 Lund, Sweden.
Abstract:A new type of amperometric biosensor based on immobilised acetylcholine esterase was designed and constructed. The enzyme was immobilised on a flow-through working electrode, which was prepared from reticulated vitreous carbon (RVC) or from a composite material consisting of RVC and superporous agarose. The sensor was operated in FIA mode using acetylthiocholine as a substrate. The sensor responded to inhibitors such as paraoxon-10(-9) mol was detected by the sensor in a non-optimised configuration. The practical lifetime of the sensor was at least 1 month.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏