首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Optimization of performance and minimization of silicate interference in continuous flow phosphate analysis
Authors:Zhang J Z  Fischer C J  Ortner P B
Institution:CIMAS, Rosenstiel School of Marine and Atmospheric Science, University of Miami, 4600 Rickenbacker Causeway, Miami FL 33149, USA; Ocean Chemistry Division, Atlantic Oceanographic and Meteorological Laboratory, National Oceanic and Atmospheric Administration, Miami, FL 33149, USA.
Abstract:Specific reaction conditions for automated continuous flow analysis of phosphate are optimized in regard to minimizing coating and silicate interference, while maintaining high sensitivity. Use of Sb in the reagent increases sensitivity and yields absorbances with little temperature dependence. Coating can be minimized by using a final solution at a pH>0.5. At final pH of 0.78 there is maximum interference from silicate in the sample. We recommend therefore as an optimal reaction condition with minimal silicate interference, the use of Sb, a final solution pH of 1.00, room temperature for the reaction and a H(+)]/Mo] ratio of 70. An equation is provided to correct silicate interference in high precision phosphate determination.
Keywords:
本文献已被 PubMed 等数据库收录!