Comparison of some analytical performance characteristics in inductively coupled plasma spectrometry of platinum group metals and gold
Authors:Kanický V  Otruba V  Mermet J M
Affiliation:Department of Analytical Chemistry, Masaryk University, CZ 611 37 Brno, Czech Republic.
Abstract:The limits of detection, precision and matrix effects in the inductively coupled plasma spectrometry of platinum group metals (PGMs) and gold were measured and evaluated for four ICP-AES and one ICP-MS instrument. The sample matrix was a cationic surfactant used for the PGMs and gold preconcentration on a modified silica gel (C18). A sorption of ion associates of PGMs and gold chlorocomplexes with the cation of onium salt N(1-carbaethoxypentadecyl)-trimethyl ammonium bromide was considered. The calibration curves, limits of detection and matrix effects were evaluated in the presence of 0.003 mol dm(-3) of onium salt (1.3 mg cm(-3)) and 0.1 mol dm(-3) HCl. The values of limits of detection (3 sigma(bl)) of PGMs for all axial ICP instruments were mostly below 10 ng cm(-3). Lateral observation on dual view ICP instrument yielded only 3 times higher detection limits in comparison to the axial mode of the same spectrometer and the detection limits for ICP-MS instrument were on the levels of units or tens of pg cm(-3). These limits of detection did not significantly differ from values obtained with pure solutions. Matrix effects in the presence of onium salt did not exceed 12% depression in the analytical signals. Besides the coefficients of correlation, the uncertainties on centroids of concentrations were calculated for calibration graphs obtained by linear regression.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号