Gamma peak search and peak fitting algorithm for a low-resolution detector with applications in gamma spectroscopy |
| |
Authors: | Lam Kenneth Zhang Weihua |
| |
Affiliation: | 1.Radiation Protection Bureau, Health Canada, 775 Brookfield Road, Ottawa, ON, K1A 1C1, Canada ; |
| |
Abstract: | A Savitzky–Golay filtering for smoothing and peak search written in Python is presented in this paper alongside its applications in the list-mode digital data acquisition dual gamma–gamma coincidence bismuth germanate (BGO) detector. The study has demonstrated that the software provides a reliable and effective way to quantify trace amounts of 22Na and 7Be in aerosol samples collected at Resolute Bay, Canada with a critical limit of 3 mBq and 5 Bq respectively for a 20 h counting interval, which are believed to be the inherent limitations of the dual-BGO system. |
| |
Keywords: | |