首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Gamma peak search and peak fitting algorithm for a low-resolution detector with applications in gamma spectroscopy
Authors:Lam  Kenneth  Zhang  Weihua
Institution:1.Radiation Protection Bureau, Health Canada, 775 Brookfield Road, Ottawa, ON, K1A 1C1, Canada
;
Abstract:

A Savitzky–Golay filtering for smoothing and peak search written in Python is presented in this paper alongside its applications in the list-mode digital data acquisition dual gamma–gamma coincidence bismuth germanate (BGO) detector. The study has demonstrated that the software provides a reliable and effective way to quantify trace amounts of 22Na and 7Be in aerosol samples collected at Resolute Bay, Canada with a critical limit of 3 mBq and 5 Bq respectively for a 20 h counting interval, which are believed to be the inherent limitations of the dual-BGO system.

Keywords:
本文献已被 SpringerLink 等数据库收录!