Multi-element analysis by energy-dispersive X-ray fluorescence using a simple calibration procedure, requiring only one or two standards |
| |
Authors: | P P Coetzee and K H Lieser |
| |
Institution: | (1) Fachbereich Anorganische Chemie und Kernchemie, Technische Hochschule, Hochschulstr. 4, D-6100 Darmstadt, Germany |
| |
Abstract: | Summary A reliable method for the determination of the slopes of calibration curves by interpolation is described for multi-element analysis by energy-dispersive X-ray fluorescence. The method is based on normalizing theoretically calculated curves of slope vs. atomic number, with the aid of one or two standards. It is applied to the analysis of powdered samples, using correction via the Comptonscattering peak to compensate for small matrix effects.
Multielementanalyse durch energiedispersive Röntgenfluorescenz mit einem einfachen Eichverfahren, das nur ein oder zwei Standards erfordert Zusammenfassung Eine zuverlässige Methode für die Bestimmung der Steigungen der Eichkurven für die Multielementanalyse durch energiedispersive Röntgenfluorescenz wird beschrieben. Die Methode basiert auf der Normierung theoretisch berechneter Kurven für die Abhängigkeit der Steigung von der Ordnungszahl mit Hilfe von ein oder zwei Standards. Sie wird angewendet auf die Analyse von Pulverproben unter Benutzung der Compton-Streupeak-Korrektur zum Ausgleich kleinerer Matrixeffekte.
On leave from Rand Afrikaans University, Johannesburg, Republic of South Africa |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|