首页 | 本学科首页   官方微博 | 高级检索  
     


Further development and optimization of a quartz tube atomizer for a gas chromatography — atomic absorption spectrometer system
Authors:D. S. Forsyth  S. Hayward
Affiliation:1. Food Research Division, Bureau of Chemical Safety, Food Directorate, Health Protection Branch, Health Canada, K1 OL2, Ottawa, Ontario, Canada
2. Bureau of Biostatistics and Computer Applications, Food Directorate, Health Protection Branch, Health Canada, K1A OL2, Ottawa, Ontario, Canada
Abstract:A new quartz T-tube furnace design based on commercial heater modules was evaluated for the determination of organolead and organotin compounds. Furnace operating conditions for both packed and capillary column operation were optimized by multivariate analysis. Response reproducibility ranged from 1.8 to 3.9% relative standard deviation. Calculated limits of detection were 1.4–2.0 pg for Et4Pb (as Pb) and 2.2–5.5 pg for Bu4Sn (as Sn). The limits of detection after application of Savitsky-Golay smoothing were 1.1–1.2 pg for Et4Pb (as Pb) and 2–3.7 pg for Bu4Sn (as Sn).
Keywords:
本文献已被 SpringerLink 等数据库收录!