首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of nicotinic acid in serum by high-performance liquid chromatography with fluorescence detection
Authors:Y Tsuruta  K Kohashi  S Ishida  Y Ohkura
Abstract:A sensitive method for the determination of nicotinic acid in serum is described which employs high-performance liquid chromatography with fluorescence detection. Nicotinic acid and 2-chloronicotinic acid as an internal standard in deproteinized serum are reacted with N,N'-dicyclohexyl-O-(7-methoxycoumarin-4-yl)methylisourea in acetone to give the corresponding fluorescent 4-hydroxymethyl-7-methoxycoumarin esters. The compounds are separated by reversed-phase chromatography on LiChrosorb RP-18 with isocratic elution using aqueous acetonitrile containing a small amount of sodium 1-hexanesulphonate as a mobile phase. The detection limit of nicotinic acid in serum was 0.2 nmol/ml. The method requires only 100 microliters of serum.
Keywords:
本文献已被 PubMed 等数据库收录!