Slurry sampling electrothermal atomic absorption spectrometric determination of chromium after separation/enrichment by mercaptoundecanoic acid modified gold coated TiO2 nanoparticles
Authors:Asl? Baysal  Suleyman Akman  Seda Demir  Mehmet Kahraman
Institution:aIstanbul Technical University, Department of Chemistry, Faculty of Arts and Sciences Maslak, Istanbul, Turkey;bYeditepe University, Department of Genetics and Bioengineering, Faculty of Engineering and Architecture, Kayisdagi, Istanbul, Turkey
Abstract:In this study, a novel preconcentration/separation technique based on the slurry analysis of chromium loaded on mercaptoundecanoic acid modified TiO2 core-Au shell nanoparticles prior to its determination by electrothermal atomic absorption spectrometry was described. For this purpose, at first, TiO2 nanoparticles were coated with gold shell and then modified with mercaptoundecanoic acid (MUA). Cr (III) was collected on the prepared sorbent by conventional batch technique. After separation of liquid phase, slurry of the sorbent was prepared and directly introduced into graphite furnace of atomic absorption spectrometry. By this way, all drawbacks due to elution procedure were eliminated. Optimum conditions for quantitative sorption and preparation of the slurry were investigated. The chromium in certificated sea-water and spiked drinking water was recovered in the range of 95% confidence level. The proposed technique was fast and simple as well as the risks of contamination and loss during elution were low. The limit of detection (3σ, N = 10) was 0.34 μg L− 1.
Keywords:Separation  Preconcentration  Chromium  Slurry sampling  Nanoparticle sorbent
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号