首页 | 本学科首页   官方微博 | 高级检索  
     检索      


HPLC separation and quantification of anionic surfactants using an automated on-line ion pair extraction system
Authors:Markus Schoester and Gerd Kloster
Institution:(1) Institut für Angewandte Physikalische Chemie, Forschungszentrum Jülich GmbH, W-5170 Jülich, Germany
Abstract:Summary Aliphatic anionic surfactants, like alkylsulfonates, alkyl sulfates, agr-sulfo alkanoic acids and their esters, were separated with respect to their alkyl chain length by high performance liquid chromatography (HPLC) using reversed phase (RP) columns. For sensitive and specific detection of these classes of compounds the HPLC separation was combined with a post-column ion-pair extraction system. The limit of detection is in the range of 3–30 ng. Previously reported detection systems of this kind were modified by changing the inner set-up of the phase separator and the extraction capillary and integrating a purge function, which led to a higher system stability and made automatization possible. In order to optimize the experimental set-up, the influence of the reagent concentration as well as different sizes and materials of the extraction capillary were evaluated. The calibration curve of surfactant concentrations versus fluorescence emission is discussed for sodium decylsulfonate. Since both the electrolyte concentration and the percentage of organic solvents in the mobile phase may adversely affect the detection system, different eluent compositions were tested.This article is part of the planned dissertation of Markus Schoester at the Mathematical Natural Science Faculty of the University of Düsseldorf
Keywords: