首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of interferences on measurements of 90Sr/90Y by TDCR Cherenkov counting technique
Authors:M Tayeb  X Dai  E C Corcoran  D G Kelly
Institution:1. Chalk River Laboratories, Atomic Energy of Canada Limited, Chalk River, ON, K0J 1J0, Canada
2. Chemistry & Chemical Engineering Department, Royal Military College of Canada, Kingston, ON, K7K 7B4, Canada
Abstract:A study to evaluate conditions affecting the determination of 90Sr/90Y activities in liquid samples by the triple-to-double coincidence ratio (TDCR) Cherenkov counting technique was conducted. The Cherenkov radiation produced by the 90Y beta decay was determined using a commercially available Hidex 300 SL liquid scintillation counter. The interferences of sample geometry, including sample counting vial type and volume composition, and sample colour on the TDCR were investigated. The effects of potentially interfering beta and mixed beta–gamma emitters on the TDCR Cherenkov counting of 90Sr/90Y activities were also examined. The TDCR values were used to quantify counting efficiencies of 90Y under different experimental conditions. The results demonstrated that the Cherenkov counting efficiency of 90Y is independent of sample volume and counting vial size. The effect of colour quenching was examined using yellow and brown food-grade dyes. The TDCR correction for colour quenching was found to be effective. An evaluation of counting efficiency of different beta-emitting radionuclides demonstrated that strong gamma emissions can contribute to the Cherenkov counting efficiency. Overall, measured radioactivity values deviated from reference values by ≤7.5 %, which is acceptable for screening applications in emergency situations.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号