首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Aptamer‐based thrombin assay on microfluidic platform
Authors:Fang‐yuan Chen  Zheng Wang  Peng Li  Hong‐Zhen Lian  Hong‐Yuan Chen
Institution:State Key Laboratory of Analytical Chemistry for Life Science, School of Chemistry and Chemical Engineering and Center of Materials Analysis, Nanjing University, , Nanjing, P. R. China
Abstract:A facile and sensitive aptamer‐based protocol has been developed for protein assay on microfluidic platform with fluorescence detection using an off‐chip microarray scanner. Aptamer‐functionalized magnetic beads were used to capture thrombin that binds to a second aptamer fluorescently labeled by Cy3. Experimental conditions, such as incubation time and temperature, washing time, interfering proteins, and aptamer, etc., were optimized for the microchip method. This work demonstrated there was a good relationship between fluorescence intensity and thrombin concentration in the range of 65–1000 ng/mL with the RSD less than 8%. Notably, an analysis only needs 1 μL volume of sample injection and this system can capture extremely tiny amount thrombin (0.4 fmol). This method has been successfully applied to assay of thrombin in human serum with the recovery of 79.74–95.94%.
Keywords:Aptamer  Magnetic beads  Microfluidic chip  Microarray scanner  Thrombin
设为首页 | 免责声明 | 关于勤云 | 加入收藏