Silver based mercury film electrode: Part IV. Comparison of theoretical and experimental voltammetric results obtained for cadmium |
| |
Authors: | Piotr Ostapczuk Zenon Kublik |
| |
Institution: | Institute of Fundamental Problems of Chemistry, University of Warsaw, Pasteura 1, 02093 Warsaw (Poland) |
| |
Abstract: | Using the silver based mercury film electrode (SBMFE) and cyclic and stripping voltammetric techniques the behaviour of cadmium has been compared with the behaviour of lead as well as with the predictions of the theory of de Vries and van Dalen. At the SBMFE with film thickness between 0.1 and 2 μm lead behaves with good agreement with theoretical predictions and only at thicknesses higher than 2 μm some deviations occur due to collecting of the excess mercury at the bottom of the wire electrode. On the other hand even at thin film electrodes the behaviour of cadmium deviates significantly from the predictions of the theory. The height of the anodic peak decreases and its width increases; also the displacement of the potentials of both cathodic and anodic peaks is smaller than the values predicted theoretically. The deviations are caused neither by the formation of intermetallic compounds in the bulk of mercury phase nor by the formation of heterogenous cadmium amalgam; they reflect the interaction between cadmium dissolved in mercury and the solid silver amalgam which is the substrate of the mercury film. Owing to the effect discussed the stripping determination of cadmium at SBMFE is characterized by a lower sensitivity and reproducibility. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |