首页 | 本学科首页   官方微博 | 高级检索  
     


Interlaboratory evaluation of hydrophone sensitivity calibration from 0.1 to 2 MHz via time delay spectrometry
Authors:Harris Gerald R  Gammell Paul M  Lewin Peter A  Radulescu Emil G
Affiliation:US Food and Drug Administration, Center for Devices and Radiological Health, 9200 Corporate Boulevard, Mail Code HFZ-132, Rockville, MD 20850, USA. grh@cdrh.fda.gov
Abstract:Knowing the low-frequency response of hydrophones, down to 100 kHz at least, is important for accurate biomedical ultrasound measurements. However, current international standards do not extend below 500 kHz. Furthermore, commercial hydrophone sources typically do not supply sensitivity data below 1-2 MHz. Therefore, to help identify and validate practical calibration methods below 2 MHz, the authors have extended their previous individual efforts in an interlaboratory evaluation of sensitivity calibration using the swept-frequency technique, time delay spectrometry (TDS). Calibrations were performed for needle and membrane PVDF hydrophones using each laboratory's TDS system. Each site employed the same purpose-built broadband source transducers, comprising both plano-concave and biconcave 1-3 piezocomposite elements 4 cm in diameter, with maximum and minimum thicknesses of approximately 1.5 and 0.1 cm. Agreement between laboratories was within the estimated measurement precision of +/-0.6 dB. The results demonstrated that a TDS system employing such transducers constitutes a viable method for hydrophone calibrations in this frequency range.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号