Installation of a Shielded SIMS for the Analysis of Irradiated Nuclear Fuels |
| |
Authors: | Stéphane Brémier Rachid Hasnaoui Stéphane Portier Olivier Bildstein Clive T Walker |
| |
Institution: | (1) The European Commission, Joint Research Centre, Institute for Transuranium Elements, Postfach 2340, D-76125 Karlsruhe, Germany |
| |
Abstract: | A CAMECA IMS 6F secondary ion mass spectrometer (SIMS) for the analysis of irradiated nuclear fuel has been installed in the
Microbeam Analysis Laboratory of the Institute for Transuranium Elements (ITU). This device is specially equipped with heavy
metal shielding to enable the safe examination of irradiated nuclear fuel samples with activities up to 75 GBq. At ITU the
shielded SIMS will be used in conjunction with EPMA taking advantage of the complementary nature of the two techniques and
will make important contributions to ongoing research programmes such as the safety of nuclear fuels, the partitioning and
transmutation programme and the characterisation of spent fuels. The paper describes the shielded SIMS installation and presents
a selection of results from the commissioning tests. |
| |
Keywords: | : Shielded SIMS irradiated nuclear materials oxidised Zircaloy Lithium Boron |