Determination of lead in wine by hydride generation atomic fluorescence spectrometry in the presence of hexacyanoferrate(III) |
| |
Authors: | Irina B. Karadjova Leonardo Lampugnani Alessandro D’Ulivo Massimo Onor Dimiter L. Tsalev |
| |
Affiliation: | (1) Faculty of Chemistry, University of Sofia, 1 James Bourchier Blvd., 1164 Sofia, Bulgaria;(2) Istituto per i Processi Chimico-Fisici (IPCF-CNR), Area della Ricerca di Pisa, Via Moruzzi 1, 56124 Pisa, Italy |
| |
Abstract: | A rapid, accurate, and precise method is described for the determination of Pb in wine using continuous-flow hydride generation atomic fluorescence spectrometry (CF-HGAFS). Sample pretreatment consists of ten-fold dilution of wine followed by direct plumbane generation in the presence of 0.1 mol L−1 HCl and 1% m/v K3[Fe(CN)6] with 1% m/v NaBH4 as reducing agent. An aqueous standard calibration curve is recommended for Pb quantification in wine sample. The method provides a limit of detection and a limit of quantification of 0.3 μg L−1 and 1 μg L−1, respectively. The relative standard deviation varies between 2–6% (within-run) and 4–11% (between-run) at 3–30 μg L−1 Pb levels in wine. Good agreement has been demonstrated between results obtained by CF-HGAFS and direct electrothermal atomic absorption spectrometry in analyses of red and white wines within the concentration range of 9.2–25.8 μg L−1 Pb. |
| |
Keywords: | Lead Wine Hydride generation atomic fluorescence spectrometry Hexacyanoferrate (III) |
本文献已被 PubMed SpringerLink 等数据库收录! |
|