首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of lead in wine by hydride generation atomic fluorescence spectrometry in the presence of hexacyanoferrate(III)
Authors:Irina B Karadjova  Leonardo Lampugnani  Alessandro D’Ulivo  Massimo Onor  Dimiter L Tsalev
Institution:(1) Faculty of Chemistry, University of Sofia, 1 James Bourchier Blvd., 1164 Sofia, Bulgaria;(2) Istituto per i Processi Chimico-Fisici (IPCF-CNR), Area della Ricerca di Pisa, Via Moruzzi 1, 56124 Pisa, Italy
Abstract:A rapid, accurate, and precise method is described for the determination of Pb in wine using continuous-flow hydride generation atomic fluorescence spectrometry (CF-HGAFS). Sample pretreatment consists of ten-fold dilution of wine followed by direct plumbane generation in the presence of 0.1 mol L−1 HCl and 1% m/v K3Fe(CN)6] with 1% m/v NaBH4 as reducing agent. An aqueous standard calibration curve is recommended for Pb quantification in wine sample. The method provides a limit of detection and a limit of quantification of 0.3 μg L−1 and 1 μg L−1, respectively. The relative standard deviation varies between 2–6% (within-run) and 4–11% (between-run) at 3–30 μg L−1 Pb levels in wine. Good agreement has been demonstrated between results obtained by CF-HGAFS and direct electrothermal atomic absorption spectrometry in analyses of red and white wines within the concentration range of 9.2–25.8 μg L−1 Pb.
Keywords:Lead  Wine  Hydride generation atomic fluorescence spectrometry  Hexacyanoferrate (III)
本文献已被 PubMed SpringerLink 等数据库收录!