A method for accurate in silico modeling of ultrasound transducer arrays
Authors:Drake A Guenther  William F Walker
Institution:1. Division of Urologic Surgery, Duke University, Durham, North Carolina;2. Department of Mechanical Engineering and Materials Science, Duke University, Durham, North Carolina;3. Department of Urology, Johannes Gutenberg University, Mainz, Germany;1. Department of Health Science and Technology, Aalborg University, Aalborg, Denmark;2. Duke Biomedical Engineering Department, Duke University, Durham, North Carolina, USA;3. Duke Department of Medicine, Duke University Hospital, Durham, North Carolina, USA;4. Department of Cardiology, Aalborg University Hospital, Aalborg, Denmark;1. Department of Internal Medicine 1, J.W. Goethe-University Hospital, Frankfurt, Germany;2. Department of Internal Medicine 3, J.W. Goethe-University Hospital, Frankfurt, Germany;1. Institute of Physical Chemistry, University of Heidelberg, D-69120 Heidelberg, Germany;2. General Physics Institute, Russian Academy of Sciences, 119991 Moscow, Russian Federation
Abstract:This paper presents a new approach to improve the in silico modeling of ultrasound transducer arrays. While current simulation tools accurately predict the theoretical element spatio-temporal pressure response, transducers do not always behave as theorized. In practice, using the probe’s physical dimensions and published specifications in silico, often results in unsatisfactory agreement between simulation and experiment. We describe a general optimization procedure used to maximize the correlation between the observed and simulated spatio-temporal response of a pulsed single element in a commercial ultrasound probe. A linear systems approach is employed to model element angular sensitivity, lens effects, and diffraction phenomena. A numerical deconvolution method is described to characterize the intrinsic electro-mechanical impulse response of the element. Once the response of the element and optimal element characteristics are known, prediction of the pressure response for arbitrary apertures and excitation signals is performed through direct convolution using available tools. We achieve a correlation of 0.846 between the experimental emitted waveform and simulated waveform when using the probe’s physical specifications in silico. A far superior correlation of 0.988 is achieved when using the optimized in silico model. Electronic noise appears to be the main effect preventing the realization of higher correlation coefficients. More accurate in silico modeling will improve the evaluation and design of ultrasound transducers as well as aid in the development of sophisticated beamforming strategies.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号