首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Plate form three-dimensional gradient coils for L-band ESR imaging experiment
Authors:Ke Wu  Changgang Huang  Yuanlin Cao  Yingguang Zheng  Jianbo Cong  Hong Xian  Changzhen Wang  Shangkai Gao  Baolu Zhao
Institution:Beijing Institute of Radiation Medicine, Beijing 100850, China.
Abstract:A set of plate form three-dimensional magnetic gradient coils was developed and used in electron spin resonance imaging (ESRI) experiment. The coils were processed with whole copper plates instead of wound with copper wires, which made its structure so compact that it was much thinner and smaller comparing to those traditionally used in ESRI. The coil set had a pie-like appearance of which the total thickness was only 14 mm and the outer diameter was 250 mm. The efficiency of the coils could be greater than 10 mT/m/A when distance between the two side-pieces was 63 mm. A maximum gradient strength of more than 200 mT/m could be obtained with driving current of about 20 A in each dimension coil. The spatial linearity was better than 5% in all three dimensions within the available spatial linearity area of larger than a sphere of 40 mm in diameter. The stability of the gradients strength could reach the level of 10(-5). An imaging resolution of better than 1 mm could be achieved with the coil set. Some preliminary practical imaging results show that the developed gradient coil set is suitable for L-band ESRI experiment of biological samples or even in vivo small animals.
Keywords:Gradient coil  Plate form  Three-dimension  ESRI
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏