Determination of hydroxyproline in meat products by flow injection analysis |
| |
Authors: | J. M?ller and A. Sj?din |
| |
Affiliation: | (1) Tecator AB, Box 70, S-263 01 Höganäs, Sweden |
| |
Abstract: | Summary A modification of the German standard method for the determination of hydroxyproline (DIN 10144) is described. By applying a thermostatted reaction coil the relatively slow developing colour reaction between pyrrols and 4-dimethylaminobenzaldehyde could be adapted to FIA-conditions. The presented FIA-method allows for an injection frequency of 80/h and a sample throughput, on routine basis, of about 40/h. The reproducibility of multiple injections in the range 20–200 mg hydroxyproline/l hydrolysate is 1.5% r.s.d. Sulfuric and hydrochloric acid have been used for hydrolysis.
Bestimmung des Hydroxyprolingehaltes von Fleischerzeugnissen mit Hilfe der Fließ-Injektionsanalyse |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|