首页 | 本学科首页   官方微博 | 高级检索  
     


Flow injection determination of iodide ion in a photographic developing solution using iodide ion-selective electrode detector
Authors:Masadome T  Sonoda R  Asano Y
Affiliation:

a Department of Chemical Science and Engineering, Ariake National College of Technology, Omuta, Fukuoka 836-8585, Japan

b Laboratory of Chemistry, Department of General Education, Ariake National College of Technology, Omuta, Fukuoka 836-8585, Japan

Abstract:A potentiometric flow injection determination method for iodide ion in a photographic developing solution was proposed by utilizing a flow-through type iodide ion-selective electrode detector. The sensing membrane of the electrode was Ag2S–AgI membrane. The response of the electrode detector as a peak-shape signal was obtained for injected iodide ion in a photographic developing solution. A linear relationship in the subnernstian zone was found to exist between peak height and the concentration of the iodide ion in a photographic developing solution in a concentration range from 0 to 6.0×10−5 mol l−1. The relative standard deviation for ten injections of 2×10−5 mol l−1 iodide ion in a photographic developing solution was 0.96% and the sampling rate was approximately 12–13 samples h−1. The iodide ion could be determined under coexisting of an organic reducing reagent and inorganic electrolytes of high concentration in a photographic developing solution sample solution by the present method.
Keywords:Photographic developing solution   Iodide-selective electrode detector   Iodide ion   Flow injection analysis
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号