Automated titration of sub-microgram amounts of metal ions in aqueous solutions |
| |
Authors: | W. J. van Oort V. W. J. van den Bergen B. Griepink |
| |
Affiliation: | (1) Laboratory for Analytical Chemistry, the University, Utrecht, The Netherlands |
| |
Abstract: | Summary Using the same simple ion selective electrode of the copper sulphide-type, about 0.2–10 nMole of Cu(II), Zn(II), Cd(II) and Pb(II) ions can be chelometrically titrated. The sample to be analyzed disturbs an equilibrium situation in a solution containing about 2–5 · 10–6 M of Cu2+-ions. In the titration of Zn, Cd and Pb the solution in equilibrium contains moreover about 0.05 mole/l of the copper-EDTA complex. The initial equilibrium signal is reset with EDTA. The standard deviation is about 0.5–2 ng (5 detn.). A titration takes about 1–4 min.
Automatische Titration von Sub-Mikrogramm-Mengen von Metallionen in wäriger LösungI. Chelatometrische Titration von Cu(II), Zn(II), Cd(II) und Pb(II) mit einer einfachen ionenselektiven Elektrode Zusammenfassung Mit einer ionenselektiven Elektrode vom CuS-Typ sind etwa 0,2–10 nMol Cu(II), Zn(II), Cd(II) and Pb(II) komplexometrisch zu bestimmen. Die Probe zerstört eine Gleichgewichtssituation in einer Lösung, die etwa 2–5 · 10–6 M Cu2+ enthält. Bei der Titration von Zn, Cd, und Pb enthält die Lösung zusätzlich noch 0,05 Mol/l des Kupferkomplexonkomplexes. Nach Zufügung der Probe wird die Anfangslage mit ÄDTA wieder hergestellt. Die Standardabweichung (5 Bestimmungen) beträgt etwa 0,5–2 ng, die Titrationsdauer 1–4 min. |
| |
Keywords: | Best. von Kupfer, Zink, Cadmium, Blei Volumetrie, Elektroden, ionenselektive Ä DTA, CuS-Typ-Elektrode |
本文献已被 SpringerLink 等数据库收录! |
|