首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Slurry sampling electrothermal atomic absorption spectrometry for trace element analysis of high-purity tungsten trioxide
Authors:V Krivan  A Heger and S Hauptkorn
Institution:(1) Sektion Analytik und H?chstreinigung, Universit?t Ulm, D-89069 Ulm, Germany, DE
Abstract:A slurry sampling ETAAS method for the determination of trace concentrations of Ca, Co, Cr, Cu, Fe, K, Mg, Na and Ni in tungsten trioxide is presented. A high background absorption appearing with large sample amounts during atomization at high temperatures was the only matrix interference observed. It could be significantly reduced by volatilization of the matrix in the pyrolysis step. Calibration curves recorded with aqueous standards could be used for calibration in all cases excluding the determination of very low analyte concentrations of Co and Fe which required the use of the standard additions method. The results obtained by this method showed an excellent agreement with those determined in digests by ETAAS. The detection limits achievable by the slurry technique were between 1 (Ca, K, Mg, Na) and 210 (Cu) ng/g. They were mostly by more than one order of magnitude better than those of the solution techniques. Received: 9 July 1997 / Revised: 6 October 1997 / Accepted: 10 October 1997
Keywords:
本文献已被 SpringerLink 等数据库收录!